Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow

986 Aufrufe

Zitiert von 7

09:41 Min.

5. März 2020

10.3791/60610-v

5. März 2020

986 Aufrufe
 ,  ,  ,  ,  ,  ,  ,  , 

This protocol describes a comprehensive hemocompatibility evaluation of blood-contacting devices using laser-cut neurovascular implants. A flow loop model with fresh, heparinized human blood is applied to mimic blood flow. After perfusion, various hematologic markers are analyzed and compared to the values gained directly after blood collection for hemocompatibility evaluation of the tested devices.

Weitere Videos entdecken

ISO 10993-4

Chapters in this video

0:05

Introduction

0:50

Heparin-Loaded Monovette Preparation and Blood Sampling

1:44

Flow Loop Preparation

2:24

Hemocompatibility Testing

2:51

Whole Blood Count Analysis

3:25

Citrate Plasma Collection

4:05

EDTA and Citrate, Theophylline, Adenosine, and Dipyridamole (CTAD) Plasma Collection

5:19

Thrombin-Anti-Thrombin III (TAT) Quantification

6:48

Scanning Electron Microscopy (SEM) Sample Preparation

7:30

Results: Representative Hemocompatibility Testing

9:02

Conclusion

Related Videos